Accessories and other models and sizes are also available. Please contact our product manager for more information.
Temperature
Temperature sensor: digital temperature sensor, mounted on the outside of the bath
Working temperature range: at least 5 above ambient to +100 (with cover, depending on ambient pressure)
Setting temperature range: +10 to +100 °C
Setting accuracy temperature: 0.1 °C
Control technology
Calibration: 2-point calibration
Controller intuitive: 3.5" colour graphic display with touch screen
Controller: digital display of set and actual temperature and of (remaining) programme time
Safety
Temperature control: two-stage safety system to prevent overtemperature
Temperature control: individually adjustable overtemperature alarm, visual and acoustic
Temperature control: Independent temperature limiter to switch off the heating in case of failure
Autodiagnostic system: microprocessor PID-temperature controllerwith integrated autodiagnostic system withfault indicator
Heating concept
Heating: corrosion-proof large-area heating on the outside of the interior
Standard equipment
Cover: Stainless steel gable cover
Drain system: innovative drain system - easy to empty
Timer: digital timer from 1 min. up to 99 hours 59 min., visual and acoustic
Stainless steel interior
Interior: Deep-drawn stainless steel tray (stainless steel W.St. 1-4301), without corners and edges for easy cleaning incl. central drain (no disturbing installations)
Dimensions: W x H x D 277 x 150 x 217 mm
Volume: 7 l
Liquid level min: 40 mm
Liquid level max: 114 mm
Textured stainless steel casing
Dimensions: 349 x 348 x 289 mm
Electrical data
Voltage: 230 V, 50/60 Hz
Electrical load: approx. 1000 W
Voltage: 115 V, 50/60 Hz
Electrical load: approx. 1000 W
Ambient conditions
Installation: The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm.
Ambient temperature: +5 °C to +40 °C
Humidity rh: max. 80 %, non-condensing
Overvoltage category: II
Pollution degree: 2
Packing/shipping data
Transport information: The appliances must be transported upright
Customs tariff number: 8419 8998
Country of origin: Federal Republic of Germany
WEEE-Reg.-No.: DE 66812464
Dimensions approx incl. carton: w x h x d: 570 x 450 x 420 mm
Net weight: approx. 6 kg